DETERMINATION OF COPPER, CADMIUM, LEAD AND BISMUTH IN HIGH-PURITY ZINC METAL SAMPLES BY ATOMIC-ABSORPTION SPECTROMETRY AFTER PRECONCENTRATION USING AMBERLITE XAD-1180 RESIN


Creative Commons License

TOKALIOGLU S., KARTAL S., Elçi L.

ANALYTICAL SCIENCES, cilt.10, sa.5, ss.779-782, 1994 (SCI-Expanded) identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 10 Sayı: 5
  • Basım Tarihi: 1994
  • Doi Numarası: 10.2116/analsci.10.779
  • Dergi Adı: ANALYTICAL SCIENCES
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Sayfa Sayıları: ss.779-782
  • Erciyes Üniversitesi Adresli: Evet

Özet

A column filled with Amberlite XAD-1180 resin was examined in order to test its suitability for extracting inorganic complexes of Cu, Cd, Pb, Bi, Ni, Fe, Mn and Co from sulfuric, nitric or hydrochloric acid solutions containing iodide, thiocyanate or their mixtures. The proposed procedure was applied to the determination of Cu, Cd, Pb and Bi in pure zinc metals by flame- or graphite furnace-atomic absorption spectrometry with good results, such as recoveries greater than or equal to 95%, relative standard deviations less than or equal to 6% (n=10) and detection limits less than or equal to 0.200 mu g cm(-3).

A  column  filled  with  Amberlite  XAD-1  l80  resin  was  examined  in  order  to  test  its  suitability  for  extracting  inorganic complexes  of  Cu,  Cd,  Pb,  Bi,  Ni,  Fe,  Mn  and  Co  from  sulfuric,  nitric  or  hydrochloric  acid  solutions  containing  iodide, 
thiocyanate  or  their  mixtures.  The  proposed  procedure  was  applied  to  the  determination  of  Cu,  Cd,  Pb  and  Bi in  pure zinc  metals  by  flame-  or  graphite  furnace-atomic  absorption  spectrometry  with  good  results,  such  as  recoveries >95%, relative  standard  deviations  <6%  (n=10)  and  detection  limits  <0.200  µg cm-3.